Troponin T - high sensitivity
9.32
Heparin (light green), SST (gold), "T-" labs SST (gold)
0.2 mL Heparin (light green) or Plain(red).
Biochemistry
Mon - Sun, 24hrs
3 hrs
Waikato - Cobas 8100 ECLIA
ng/L
< 15 ng/L
Waikato - 5%
The analyser at the main Laboratory changed from a e601 to a e801 on 26 August 2020. Results increased by approximately 8% and decreased by 1 ng/L at the low end.
There are multiple causes of elevated troponin levels. The diagnosis of a Type 1 myocardial infarction requires a setting consistent with ischaemia (symptoms and/or ECG changes) and hsTnT >= 15 ng/L with a rise or fall of 50% if hsTnT is 15-50 ng/L or a rise or fall of 20% if hsTnT is greater than 50 ng/L (as per cardiology algorithm).
0.1 mL.
TROPTHS
$TROPTHS
6598-7