× Test Guide Collection Guides Guidelines News Contact Useful Links

Laboratory Test Reference Guide

Laboratory Information

Test Name

pCO2

External Price (excl. GST)

11.67

Specimen Collection

Specimen Collection and Transport Protocols

Remove at least 3 ml of blood to waste prior to collecting the blood sample from a line.  Must be analysed within 15 minutes of collection.

Adult Specimen

1.0 ml blood in an adult blood gas syringe or 0.5 ml in a paediatric blood gas syringe

Paediatric Specimen

0.5 ml blood in a paediatric blood gas syringe

Sample Delivery to Lab

Ambient

Test Information

Department

Biochemistry and POCT

Test Availability

Mon - Sun, 24hrs

Laboratory Turnaround Time

30 minutes

Method

Blood gas analyser

Unit of measurement

kPa

Reference interval

Arterial 4.6 - 6.0 kPa       Venous 5.0 - 6.4 kPa

Cord Blood Arterial 5.6 - 9.8 kPa       Cord Blood Venous 3.8 - 7.1 kPa

Uncertainty of measurement

± 5%

Minimum Volume

0.3 ml collected into a paediatric blood gas syringe

Electronic Information

Registration Code

PCO2

HL7 mapping code

$PCO2

LOINC Code

11557-6